Airway Management
Hand Held Automatic Resuscitators
Automatic Transport Ventilators
Bag-Valve-Mask Resuscitators
Carrying Cases
Cylinders & Facemasks
CPR Protective Barrier Devices
Demand Valves
Emergency Burn Management
Oxygen Therapy
Regulators
Training - Ventilation



 
 
CAREvent® EMT  
   
   
Download PDF
spec. sheet
View product comparison chart
 
The “NEW” CAREvent® EMT - Automatic Resuscitation with A.T.T.I.T.U.D.E.®!

Trained individuals now have a safe, effective and inexpensive means of providing positive pressure ventilation during respiratory and/or cardiac arrest.

Providing “TRUE AUTOMATIC VENTILATION” at an affordable price, the lightweight, portable, simple to use and extremely durable CAREvent EMT® Handheld Resuscitator is designed for the demands of the emergency resuscitation and transport environment. In addition, this pneumatically powered, time/volume cycled ventilator has the added feature of a Manually Actuated, Automatic Ventilation Override Button (Manual Button) for use during CPR.

Depressing the Manual Button overrides the automatic cycling and allows the operator to provide the breaths manually in time with the chest compressions. Depression of the button inhibits the automatic cycling for 10 seconds. This allows the rescuer to provide the 15 chest compressions in accordance with the Guidelines 2000 for Resuscitation (G2K®) without the ventilator attempting to provide the next breath during the compression cycle. If the Manual Button is not depressed again the automatic cycling will re-start without any intervention by the operator. The second breath can then be provided manually. The ventilator cycling will then stop for 10 seconds, allowing the next set of compressions to be provided.

The Single Slider Control provides the simultaneous adjustment of ventilation frequency and tidal volume. The CAREvent EMT® has six, preset, tidal volume / frequency settings for a range of patients from 15 kg., body weight to large adults. An OFF position is provided so that the device can be left attached to the gas supply with the supply turned on ready for immediate use.

The patient is protected from barotrauma due to over inflation by the Pressure Relief System (with audible alarm). This warns the operator of airway obstruction or restriction. The excess delivered volume is discharged to atmosphere when the alarm activates.

Features  

  • A.utomatic
  • T.ime cycled
  • T.idal volume/frequency controlled
  • I.:E ratio constant (1:2)
  • T.rigger (MANUAL OVERRIDE) for CPR
  • U.naffected by lung/airway changes
  • D.elay on Manual Override - G2K* compliant.
  • E.xtremely low drive gas consumption

    Specifications  

      Tidal Volume:
    0.15 - 0.6 litres
      Breaths Per Minute: 20 - 10
      I:E Ratio: 1:2
      Manual Flow Rate Range: As per Vt/BPM Control Setting
      Input Pressure: 50 PSI (+/- 10 PSI)
      Audible Pressure Relief Valve: 60 cm H20
      Operating Temperature: -18°C to +50°C
    (0°F to 122°F)
      Storage Temperature: -40°C to +60°C
    (-40°F to 140°F)
      Relative Humidity for Storage and Operating Use:
    15 to 95%
      Input Connection: Fixed
      Hose connection to regulator: 9/16 DISS
      Patient Valve Dead Space: 8 ml.
      Dimensions (inches): 5.5 x 2.5 x 2.9 (approx.)
      Dimensions (millimeters): 140 x 63 x 73 (approx.)
      Weight: 0.95lbs / 0.43kg
     
    (All specifications are subject to a tolerance of +/- 10%
    except the I:E Ratio which is subject to a tolerance of +/- 20%)

    Part Number and Product Description  

      01CV0500 CAREvent EMT® Handheld Automatic Resuscitator and Supply Hose, c/w 5 Single-Use Patient Valves with Universal Resuscitation Facemasks.
     
    CAREvent® BLS  
       
       
    Download PDF
    spec. sheet
    View product comparison chart
    Download video
     

    The CAREvent® BLS Handheld Resuscitator is lightweight, portable, extremely durable and simple to use automatic time/volume cycled ventilatory resuscitator. The added feature of a Manually Actuated, Automatic Ventilation Override Button (Manual Button) allows the operator to control the ventilations manually if required.

    The single slider control for the simultaneous adjustment of ventilation frequency and tidal volume has 6 preset settings for a range of patients from infants to large adults and an OFF position so that the resuscitator can be left attached to the gas supply with the supply turned on, ready for immediate use.

    During CPR actuation of the manual button causes the automatic cycling to cease to allow the provision of chest compressions. Cycling will re-start following manual actuation after a delay of 4 to 7 seconds depending on the selected setting of the slider control.

    The pneumatic logic circuit can be run off either compressed air or medical oxygen.

    CAREvent® BLS “G2K” Resuscitator

    Designed to comply with the latest CPR Guidelines 2000 recommendations of a 15:2 compression:ventilation ratio in the un-intubated patient, this model has a 10 second delay to automatic cycling re-start “in any setting”, following manual ventilation actuation, to allow for 15 chest compressions to be provided.

    Features  

  • Totally pneumatic operation
  • Lightweight
  • Durable
  • Automatic & Manual Cycling
  • Preset Pressure Relief
  • G2K Compliant Version Available
  • 2 Year Warranty Against Manufacturers Defects

    Specifications  

      TIDAL VOLUME:
    0.2-1.1 litres
      BREATHS PER MINUTE: 20-12
      I:E RATIO 1:2
      AUTOMATIC FLOW RATE: 12-39.6 LPM
      MANUAL FLOW RATE: As per selected setting
      MINIMUM INPUT PRESSURE: 45 PSI
      MAXIMUM INPUT PRESSURE: 70 PSI
      MAX . AIRWAY PRESSURE: 60 cm H2O
      OPERATING TEMPERATURE: -18°C to +50°C
    0°F to +122°F
      STORAGE TEMPERATURE: -40°C to +60°C
    -40°F to + 140°F
      PATIENT CONNECTION: 9/16" DISS
      CYLINDER DURATION: (a) 20 BPM / 0.2 I VT 100 minutes (approx.)
    (b) 12 BPM / 1.1 I V 30 minutes (approx.)
    (Based on Aluminum "D" size cylinder containing 415 litres of oxygen)
     
    (All specifications are subject to a tolerance of +/- 10%
    except the I:E Ratio which is subject to a tolerance of +/- 20%)

    Part Number and Product Description  

      01CV1000 CAREvent® BLS Handheld Automatic Time/volume Cycled Resuscitator c/w 6 foot supply hose and Universal Face Mask
      01CV1100 CAREvent® BLS Handheld Resuscitator kit with Adjustable 0 – 25LPM Regulator with 2 DISS Self-Sealing outlets in a moulded Carrying Case c/w 6 foot supply hose and Universal Face Mask
      01CV1112 CAREvent® BLS Handheld Resuscitator kit with Adjustable 0 – 25LPM Regulator with 2 DISS Self-Sealing outlets and STATVAC® II Aspirator in a moulded Carrying Case c/w 6 foot supply hose Yankauer Suction Tip with Tubing and Universal Face Mask
     
    CAREvent® ALS  
       
       
    Download PDF
    spec. sheet
    View product comparison chart
    Download video
     

    The CAREvent® ALS Handheld Resuscitator is lightweight, portable, extremely durable and simple to use automatic time/volume cycled ventilatory resuscitator with “Demand Breathing Capability” for the spontaneously breathing patient. The added feature of a Manually Actuated, Automatic Ventilation Override Button (Manual Button) allows the operator to control the ventilations manually if required.

    The single slider control for the simultaneous adjustment of ventilation frequency and tidal volume has 6 preset settings for a range of patients from infants to large adults and an OFF position so that the resuscitator can be left attached to the gas supply with the supply turned on, ready for immediate use.

    During CPR actuation of the manual button causes the automatic cycling to cease to allow the provision of chest compressions. Cycling will re-start following manual actuation after a delay of 4 to 7 seconds depending on the selected setting of the slider control.

    The Demand Breathing feature allows the patient to spontaneously breathe through the resuscitator on 100% oxygen, at their own desired rate and volume. In addition, the patient’s inspiratory efforts cause the automatic cycling to cease. Should the patient cease breathing spontaneously, the ventilator will automatically restart cycling in the automatic setting selected.

    Note: In the child settings the inspired flow rate of the patient may be insufficient to cause the Automatic Circuit Shutoff to function.

    The pneumatic logic circuit can be run off either compressed air or medical oxygen.

    CAREvent® ALS “G2K” Resuscitator

    Designed to comply with the latest CPR Guidelines 2000 recommendations of a 15:2 compression:ventilation ratio in the un-intubated patient, this model has a 10 second delay to automatic cycling re-start “in any setting”, following manual ventilation actuation, to allow for 15 chest compressions to be provided.

    Features  

  • Totally pneumatic operation
  • Lightweight
  • Durable
  • Automatic & Manual Cycling
  • Demand Breathing Capability with auto circuit shut off
  • Preset Pressure Relief
  • G2K Compliant Version Available
  • 2 Year Warranty Against Manufacturers Defects

    Specifications  

      TIDAL VOLUME:
    0.2-1.1 litres
      BREATHS PER MINUTE: 20-12
      I:E RATIO 1:2
      AUTOMATIC FLOW RATE: 12-39.6 LPM
      MANUAL FLOW RATE: As per selected setting
      DEMAND FLOWRATE: Minimum 120 LPM
      MINIMUM INPUT PRESSURE: 45 PSI
      MAXIMUM INPUT PRESSURE: 70 PSI
      MAX . AIRWAY PRESSURE: 60 cm H2O
      OPERATING TEMPERATURE: -18°C to +50°C
    0°F to +122°F
      STORAGE TEMPERATURE: -40°C to +60°C
    -40°F to + 140°F
      PATIENT CONNECTION: 9/16" DISS
      CYLINDER DURATION: (a) 20 BPM / 0.2 I VT 100 minutes (approx.)
    (b) 12 BPM / 1.1 I V 30 minutes (approx.)
    (Based on Aluminum "D" size cylinder containing 415 litres of oxygen)
     
    (All specifications are subject to a tolerance of +/- 10%
    except the I:E Ratio which is subject to a tolerance of +/- 20%)

    Part Number and Product Description  

      01CV3000 CAREvent® ALS Handheld Automatic Time/volume Cycled Resuscitator c/w 6 foot supply hose and Universal Face Mask
      01CV3100 CAREvent® ALS Handheld Resuscitator kit with Adjustable 0 – 25LPM Regulator with 2 DISS Self-Sealing outlets in a moulded Carrying Case c/w 6 foot supply hose and Universal Face Mask
      01CV3112 CAREvent® ALS Handheld Resuscitator kit with Adjustable 0 – 25LPM Regulator with 2 DISS Self-Sealing outlets and STATVAC® II Aspirator in a moulded Carrying Case c/w 6 foot supply hose Yankauer Suction Tip with Tubing and Universal Face Mask
     
    CAREvent® ALS+  
       
       
    Download PDF
    spec. sheet
    View product comparison chart
    Download video
     

    The CAREvent® ALS+ is lightweight, portable, extremely durable and simple to use, control module style, automatic time/volume cycled ventilatory resuscitator with “Demand Breathing Capability” for the spontaneously breathing patient. The added feature of a Manually Actuated, Automatic Ventilation Override Button (Manual Button) allows the operator to control the ventilations manually if required.

    The single rotary control provides simultaneous adjustment of ventilation frequency and tidal volume has 12 preset settings for a range of patients from infants to large adults. A Manual/Auto switch allows the resuscitator to be left attached to the gas supply with the supply turned on, ready for immediate use.

    During CPR actuation of the manual button causes the automatic cycling to cease to allow the provision of chest compressions. Cycling will re-start following manual actuation after a delay of 4 to 7 seconds depending on the selected setting of the slider control.

    The Demand Breathing feature allows the patient to spontaneously breathe through the resuscitator on 100% oxygen, at their own desired rate and volume. In addition, the patient’s inspiratory efforts cause the automatic cycling to cease. Should the patient cease breathing spontaneously, the ventilator will automatically restart cycling in the automatic setting selected.

    Note: In the child settings the inspired flow rate of the patient may be insufficient to cause the Automatic Circuit Shut off to function.

    The pneumatic logic circuit can be run off either compressed air or medical oxygen.

    Designed to comply with the latest CPR Guidelines 2000 recommendations of a 15:2 compression:ventilation ratio in the un-intubated patient, the CAREvent® ALS+ “G2K” model has a 10 second delay to automatic cycling re-start “in any setting”, following manual ventilation actuation, to allow for 15 chest compressions to be provided.

    Features  

  • Totally pneumatic operation
  • Lightweight
  • Durable
  • Control Module Style
  • Manual/Auto Selector Switch
  • Automatic & Manual Cycling
  • Demand Breathing Capability with auto circuit shut off
  • Preset Pressure Relief
  • G2K Compliant Version Available
  • 2 Year Warranty Against Manufacturers Defects

    Specifications  

      TIDAL VOLUME:
    0.2-1.35 litres
      BREATHS PER MINUTE: 20-10
      I:E RATIO: 1:2
      AUTOMATIC FLOW RATE: 12- 40 LPM
      MANUAL FLOW RATE: As per selected setting
      DEMAND FLOWRATE: Minimum 120 LPM
      MINIMUM INPUT PRESSURE: 45 PSI
      MAXIMUM INPUT PRESSURE: 70 PSI
      MAX. AIRWAY PRESSURE: 60 cm H2O
      OPERATING TEMPERATURE: -18°C to +50°C
    0°F to +122°F
      STORAGE TEMPERATURE: -40°C to +60°C
    -40°F to + 140°F
      PATIENT CONNECTION: 9/16" DISS
      CYLINDER DURATION: (a) 20 BPM / 0.2 I VT 100 minutes (approx.)
    (b) 12 BPM / 1.1 I V 30 minutes (approx.)
    (Based on Aluminum "D" size cylinder containing 415 litres of oxygen)
     
    (All specifications are subject to a tolerance of +/- 10%
    except the I:E Ratio which is subject to a tolerance of +/- 20%)

    Part Number and Product Description  

      01CV4000 CAREvent® ALS+ Control Module Automatic Time/volume Cycled Resuscitator c/w 6 foot supply hose, Deluxe Disposable Transport Ventilator Circuit, Test Lung and Universal Face Mask
     
    CAREvent® ATV  
       
       
    Download PDF
    spec. sheet
    View product comparison chart
    Download video
     

    The CAREvent® ATV is an Intermittent Positive Pressure ventilator with independent frequency of ventilation and Minute Volume settings to provide a comprehensive selection of delivered tidal volumes and ventilation rates. The design of the micro-pneumatic circuitry in the Automatic Transport Ventilator maintains a consistent I:E Ratio set at 1:2 throughout the entire setting range. Designed for use by suitably trained personnel in the prehospital, inter-hospital and air ambulance settings. The various modes of operation of the ventilator support the resuscitation and transportation of a wide range of patients (from infants to adults) who require ventilatory support.

    The added feature of a Manually Actuated, Automatic Ventilation Override Button (Manual Button) to allow the operator to control the ventilations manually if required. In the AUTOmatic mode, the automatic cycling will re-start following manual actuation after a delay of 4 to 7 seconds depending on the selected setting. In the MANual mode the automatic cycling is completely inhibited allowing the operator to fully control the ventilations.

    Peak airway pressure is operator adjustable to allow for differing patient conditions.

    The "Demand Breathing" feature allows spontaneously breathing patients to breathe on oxygen through the ventilator. The "Automatic Circuit Shut Off" control will cause the automatic cycling to cease when the patient being ventilated takes a spontaneous breath.

    Note: In the child settings the inspired flow rate of the patient may be insufficient to cause the Automatic Circuit Shut off to function.

    The CAREvent® ATV consists of a ruggedly constructed, portable control module, input hose and optional single use or reusable patient circuit. The ergonomically designed control groupings facilitate the selection and setting of the breathing parameters. The colour groupings provided on the controls add to this ease of use and concept.

    The unique pneumatic alarm for low input pressure improves the level of safety for the patient by warning the operator of any problems in the gas supply.

    The air mix capability of the ventilator improves the operating time on bottled oxygen for long duration transports. The extremely low drive gas consumption ensures full utilization of the cylinder contents. As an added precaution a replaceable filter is incorporated into the rear of the ventilator to prevent contaminated air entering into the unit when using the air mix mode.

    Designed to comply with the latest CPR Guidelines 2000 recommendations of a 15:2 compression:ventilation ratio in the un-intubated patient, the CAREvent® ATV “G2K” model has a 10 second delay to automatic cycling re-start “in any setting”, following manual ventilation actuation, to allow for 15 chest compressions to be provided.

    Features  

  • Totally pneumatic operation
  • Durable construction
  • Independent Minute volume and Frequency controls
  • Manual/Auto Selector Switch
  • Automatic & Manual Cycling
  • Low input Pressure Alarm
  • Demand Breathing Capability with auto circuit shut off
  • Adjustable Pressure Relief
  • G2K Compliant Version Available
  • 2 Year Warranty Against Manufacturers Defects

    Specifications  

      MINUTE VOLUME:
    2 - 14 litres
      BREATHS PER MINUTE: 8 - 40
      I:E RATIO: 1:2
      AUTOMATIC FLOW RATE: 12- 40 LPM
      MANUAL FLOW RATE: As per selected setting
      DEMAND FLOWRATE: Minimum 120 LPM
      DEMAND TRIGGERING PRESSURE -5 cm H20 @ 100 LPM
      MINIMUM INPUT PRESSURE: 45 PSI
      MAXIMUM INPUT PRESSURE:
    70 PSI
      MAX . AIRWAY PRESSURE: 60 cm H2O
      OXYGEN CONCENTRATION: 60 or 100%
      ADJUSTABLE AIRWAY PRESSURE RANGE: 20 – 60 cm H2O
      OPERATING TEMPERATURE: -18°C to +50°C
    0°F to +122°F
      STORAGE TEMPERATURE: -40°C to +60°C
    -40°F to + 140°F
      PATIENT CONNECTION: 9/16” DISS
      PATIENT VALVE DEADSPACE: 8 ml
      GAS SUPPLY STATUS INDICATOR:
    VISUAL
      LOW INPUT PRESSURE ALARM: PULSED VISUAL and AUDIBLE
      CYLINDER DURATION: AT 5L MV - 75minutes (approx.)
    (Based on Aluminum "D" size cylinder containing 415 litres of oxygen)
     
    (All specifications are subject to a tolerance of +/- 10%
    except the I:E Ratio which is subject to a tolerance of +/- 20%)

    Part Number and Product Description  

      01CV5000 CAREvent® ATV Control Automatic Transport Ventilator c/w 6 foot supply hose, Deluxe Disposable Transport Ventilator Circuit, Test Lung and Universal Face Mask
     
    CAREvent® ATV+  
       
       
    Download PDF
    spec. sheet
    View product comparison chart
    Download video
     

    The CAREvent® ATV+ is an Intermittent Positive Pressure ventilator with independent frequency of ventilation and Minute Volume settings to provide a comprehensive selection of delivered tidal volumes and ventilation rates. The design of the micro-pneumatic circuitry in the Automatic Transport Ventilator maintains a consistent I:E Ratio set at 1:2 throughout the entire setting range. Designed for use by suitably trained personnel in the prehospital, inter-hospital and air ambulance settings. The various modes of operation of the ventilator support the resuscitation and transportation of a wide range of patients (from infants to adults) who require ventilatory support.

    The added feature of a Manually Actuated, Automatic Ventilation Override Button (Manual Button) to allow the operator to control the ventilations manually if required. In the AUTOmatic mode, the automatic cycling will re-start following manual actuation after a delay of 4 to 7 seconds depending on the selected setting. In the MANual mode the automatic cycling is completely inhibited allowing the operator to fully control the ventilations.

    Peak airway pressure is operator adjustable to allow for differing patient conditions.

    The "Demand Breathing" feature allows spontaneously breathing patients to breathe on oxygen through the ventilator. The "Automatic Circuit Shut Off" control will cause the automatic cycling to cease when the patient being ventilated takes a spontaneous breath.

    Note: In the child settings the inspired flow rate of the patient may be insufficient to cause the Automatic Circuit Shut off to function.

    The CAREvent® ATV+ Automatic Transport Ventilator consists of a ruggedly constructed, portable control module, input hose and optional single use or reusable patient circuit. The ergonomically designed control groupings facilitate the selection and setting of the breathing parameters. The colour groupings provided on the controls add to this ease of use and concept.

    The unique pneumatic alarm for low input pressure improves the level of safety for the patient by warning the operator of any problems in the gas supply. An alarm for patient circuit disconnection or insufficient tidal volume delivery is also incorporated in all three models. This pneumatically powered alarm provides both a visual and audible indication of patient circuit disconnection. The silence button will override the alarm for 15 seconds.

    Should the operator not correct the problem in that time the alarm will recommence.
    The air mix capability of the ventilator improves the operating time on bottled oxygen for long duration transports. The extremely low drive gas consumption ensures full utilization of the cylinder contents. As an added precaution a replaceable filter is incorporated into the rear of the ventilator to prevent contaminated air entering into the unit when using the air mix mode.

    Designed to comply with the latest CPR Guidelines 2000 recommendations of a 15:2 compression:ventilation ratio in the un-intubated patient, the CAREvent® ATV “G2K” model has a 10 second delay to automatic cycling re-start “in any setting”, following manual ventilation actuation, to allow for 15 chest compressions to be provided.

    Features  

  • Totally pneumatic operation
  • Durable construction
  • Independent Minute volume and Frequency controls
  • Manual/Auto Selector Switch
  • Automatic & Manual Cycling
  • Circuit Disconnect Alarm
  • Low input Pressure Alarm
  • Demand Breathing Capability with auto circuit shut off
  • Adjustable Pressure Relief
  • G2K Compliant Version Available
  • 2 Year Warranty Against Manufacturers Defects

    Specifications  

      MINUTE VOLUME:
    2 - 14 litres
      BREATHS PER MINUTE: 8 - 40
      I:E RATIO: 1:2
      AUTOMATIC FLOW RATE: 12- 40 LPM
      MANUAL FLOW RATE: As per selected setting
      DEMAND FLOWRATE: Minimum 120 LPM
      DEMAND TRIGGERING PRESSURE: -5 cm H20 @ 100 LPM
      MINIMUM INPUT PRESSURE: 45 PSI
      MAXIMUM INPUT PRESSURE:
    70 PSI
      MAX . AIRWAY PRESSURE: 60 cm H2O
      OXYGEN CONCENTRATION: 60 or 100%
      ADJUSTABLE AIRWAY PRESSURE RANGE: 20 – 60 cm H2O
      OPERATING TEMPERATURE: -18°C to +50°C
    0°F to +122°F
      STORAGE TEMPERATURE: -40°C to +60°C
    -40°F to + 140°F
      PATIENT CONNECTION: 9/16” DISS
      PATIENT VALVE DEADSPACE: 8 ml
      GAS SUPPLY STATUS INDICATOR:
    VISUAL
      CIRCUIT DISCONNECT ALARM:
    PULSED AUDIBLE, AND VISUAL. ALARM SILENCE CONTROL – 15 SECONDS
      LOW INPUT PRESSURE ALARM: PULSED VISUAL and AUDIBLE
      CYLINDER DURATION: AT 5L MV - 75minutes (approx.)
    (Based on Aluminum "D" size cylinder containing 415 litres of oxygen)
     
    (All specifications are subject to a tolerance of +/- 10%
    except the I:E Ratio which is subject to a tolerance of +/- 20%)

    Part Number and Product Description  

      01CV6000 CAREvent® ATV+ Control Automatic Transport Ventilator c/w 6 foot supply hose, Deluxe Disposable Transport Ventilator Circuit, Test Lung and Universal Face Mask
     
    CAREvent® MRI  
       
       
    Download PDF
    spec. sheet
    View product comparison chart
    Download video
     

    The CAREvent® MRI is an Intermittent Positive Pressure ventilator with independent frequency of ventilation and Minute Volume settings to provide a comprehensive selection of delivered tidal volumes and ventilation rates. The design of the micro-pneumatic circuitry in the Automatic Transport Ventilator maintains a consistent I:E Ratio set at 1:2 throughout the entire setting range. Designed for use by suitably trained personnel in the prehospital, inter-hospital and air ambulance settings. The various modes of operation of the ventilator support the resuscitation and transportation of a wide range of patients (from infants to adults) who require ventilatory support.

    The added feature of a Manually Actuated, Automatic Ventilation Override Button (Manual Button) to allow the operator to control the ventilations manually if required. In the AUTOmatic mode, the automatic cycling will re-start following manual actuation after a delay of 4 to 7 seconds depending on the selected setting. In the MANual mode the automatic cycling is completely inhibited allowing the operator to fully control the ventilations.

    Peak airway pressure is operator adjustable to allow for differing patient conditions.

    The "Demand Breathing" feature allows spontaneously breathing patients to breathe on oxygen through the ventilator. The "Automatic Circuit Shut Off" control will cause the automatic cycling to cease when the patient being ventilated takes a spontaneous breath.

    Note: In the child settings the inspired flow rate of the patient may be insufficient to cause the Automatic Circuit Shut off to function.

    The CAREvent® MRI Automatic Transport Ventilator consists of a ruggedly constructed, portable control module, input hose and optional single use or reusable patient circuit. The ergonomically designed control groupings facilitate the selection and setting of the breathing parameters. The colour groupings provided on the controls add to this ease of use and concept.

    The unique pneumatic alarm for low input pressure improves the level of safety for the patient by warning the operator of any problems in the gas supply. An alarm for patient circuit disconnection or insufficient tidal volume delivery is also incorporated in all three models. This pneumatically powered alarm provides both a visual and audible indication of patient circuit disconnection. The silence button will override the alarm for 15 seconds. Should the operator not correct the problem in that time the alarm will recommence.

    The air mix capability of the ventilator improves the operating time on bottled oxygen for long duration transports. The extremely low drive gas consumption ensures full utilization of the cylinder contents. As an added precaution a replaceable filter is incorporated into the rear of the ventilator to prevent contaminated air entering into the unit when using the air mix mode.

    Designed to comply with the latest CPR Guidelines 2000 recommendations of a 15:2 compression:ventilation ratio in the un-intubated patient, the CAREvent® ATV “G2K” model has a 10 second delay to automatic cycling re-start “in any setting”, following manual ventilation actuation, to allow for 15 chest compressions to be provided.

    Features  

  • MRI compatible
  • Totally pneumatic operation
  • Durable construction
  • Independent Minute volume and Frequency controls
  • Manual/Auto Selector Switch
  • Automatic & Manual Cycling
  • Circuit Disconnect Alarm
  • Low input Pressure Alarm
  • Demand Breathing Capability with auto circuit shut off
  • Adjustable Pressure Relief
  • G2K Compliant Version Available
  • 2 Year Warranty Against Manufacturers Defects

    Specifications  

      MINUTE VOLUME: 2 - 14 litres
      BREATHS PER MINUTE: 8 - 40
      I:E RATIO: 1:2
      AUTOMATIC FLOW RATE: 12- 40 LPM
      MANUAL FLOW RATE: As per selected setting
      DEMAND FLOWRATE: Minimum 120 LPM
      DEMAND TRIGGERING PRESSURE: -5 cm H20 @ 100 LPM
      MINIMUM INPUT PRESSURE: 45 PSI
      MAXIMUM INPUT PRESSURE:
    70 PSI
      MAX . AIRWAY PRESSURE: 60 cm H2O
      OXYGEN CONCENTRATION: 60 or 100%
      ADJUSTABLE AIRWAY PRESSURE RANGE: 20 – 60 cm H2O
      OPERATING TEMPERATURE: -18°C to +50°C
    0°F to +122°F
      STORAGE TEMPERATURE: -40°C to +60°C
    -40°F to + 140°F
      PATIENT CONNECTION: 9/16” DISS
      PATIENT VALVE DEADSPACE: 8 ml
      GAS SUPPLY STATUS INDICATOR:
    VISUAL
      CIRCUIT DISCONNECT ALARM:
    PULSED AUDIBLE, AND VISUAL.
    ALARM SILENCE CONTROL – 15 SECONDS
      LOW INPUT PRESSURE ALARM: PULSED VISUAL and AUDIBLE
      CYLINDER DURATION: AT 5L MV - 75minutes (approx.)
    (Based on Aluminum "D" size cylinder containing 415 litres of oxygen)
     
    (All specifications are subject to a tolerance of +/- 10%
    except the I:E Ratio which is subject to a tolerance of +/- 20%)

    Part Number and Product Description  

      01CV7000 CAREvent® MRI Control Automatic Transport Ventilator c/w 6 foot supply hose, Deluxe Disposable Transport Ventilator Circuit, Test Lung and Universal Face Mask
     
    CAREvent® CA  
       
       
    Download PDF
    spec. sheet
    View product comparison chart
     
    Specifically designed for the “Chemical Agent Resuscitation Environment”, the CAREvent® CA Handheld Resuscitator is lightweight, portable, extremely durable and simple to use automatic time/volume cycled ventilatory resuscitator with “Demand Breathing Capability” for the spontaneously breathing patient. The added feature of a Manually Actuated, Automatic Ventilation Override Button (Manual Button) allows the operator to control the ventilations manually if required.

    The CAREvent® CA combines simplicity of operation with technological sophistication to provide automatic ventilation, at a rate and volume designed to reduce the risk of gastric insufflation. It also provides Demand Breathing and "Automatic Circuit Shut Off" of the automatic cycling unit should the patient begin adequate spontaneous respiration.

    Designed to comply with the latest CPR Guidelines 2000 recommendations of a 15:2 compression:ventilation ratio in the un-intubated patient, actuation of the manual button causes the automatic cycling to cease. This model has a 10 second delay to automatic cycling re-start “in any setting”, following manual ventilation actuation, to allow for 15 chest compressions to be provided.

    The Demand Breathing feature allows the patient to spontaneously breathe through the resuscitator on 100% oxygen, at their own desired rate and volume. In addition, the patient’s inspiratory efforts cause the automatic cycling to cease. Should the patient cease breathing spontaneously, the ventilator will automatically restart cycling in the automatic setting selected.

    The added feature of this unit is the MASK PURGE (2 lpm) CONSTANT BLEED FLOW. This reduces the risk of drawing toxic atmosphere into the mask if a leak in the mask to face seal occurs.

    Note: In the child settings the inspired flow rate of the patient may be insufficient to cause the Automatic Circuit Shut off to function.

    The pneumatic logic circuit can be run off either compressed air or medical oxygen.

    Features  

  • Totally pneumatic operation
  • Lightweight
  • Durable
  • Automatic & Manual Cycling
  • Mask purge “Bleed Flow”
  • Demand Breathing Capability with auto circuit shut off
  • Preset Pressure Relief
  • 2 Year Warranty Against Manufacturers Defects

    Specifications  

      TIDAL VOLUME:
    0.8 litres
      BREATHS PER MINUTE: 12
      I:E RATIO: 1:2
      AUTOMATIC FLOW RATE: 28.8 LPM
      MANUAL FLOW RATE: 28.8 LPM
      DEMAND FLOWRATE:
    Minimum 120 LPM
      MSK PURGE BLEED FLOW: 2 LPM
      MINIMUM INPUT PRESSURE: 45 PSI
      MAXIMUM INPUT PRESSURE:
    70 PSI
      MAX . AIRWAY PRESSURE: 60 cm H2O
      OPERATING TEMPERATURE: -18°C to +50°C
    0°F to +122°F
      STORAGE TEMPERATURE: -40°C to +60°C
    -40°F to + 140°F
      PATIENT CONNECTION: 9/16" DISS
      CYLINDER DURATION: 12 BPM / 0.8 VT 43 minutes (approx.)
    (Based on Aluminum "D" size cylinder containing 415 litres of oxygen)
     
    (All specifications are subject to a tolerance of +/- 10%
    except the I:E Ratio which is subject to a tolerance of +/- 20%)

    Part Number and Product Description  

      01CV3000-CA CAREvent® CA Chemical Agent Resuscitation Environment Handheld Automatic Time/volume Cycled Resuscitator c/w 6 foot supply hose and Universal Face Mask
     
    Genesis® II A/C  
       
       
    View product comparison chart
     
    This model is designed to meet the varying requirement of the Emergency Medical services and Hospital Personnel who may be called upon to resuscitate a wide range of patient sizes from children (over 20 kg body weight) to adults.

    All controls are located on the front of the resuscitator facing the operator including the manual button. This allows for both hands to remain on the face mask facilitating a superior mask-to-face seal during ventilation.

    The tidal volume and frequency can be simultaneously adjusted by one setting selector to suite the average patient's lung capacity and normal respiratory rates. This model is equipped with a manual button which functions independently of the automatic cycling circuit to enable the operator to control the rate and volume of ventilation. The manual flow rate is fixed at 40 lpm in line with the American Heart Association "Recommendations and Guidelines for Resuscitation".

    Spontaneously breathing patients, whose respirations are adequate, will cause the "Automatic Shut Off Circuit" to inhibit automatic ventilation. This allows the patient to breath, "on demand", at his/her own rate and volume. If the patient's respirations become inadequate, or completely cease, the automatic cycling will restart without any intervention from the rescuer.

    Features  

  • Totally pneumatic operation
  • Lightweight
  • Durable
  • Automatic & Manual Cycling
  • Demand Breathing Capability with auto circuit shut off
  • Preset Pressure Relief
  • 2 Year Warranty Against Manufacturers Defects

    Specifications  

      TIDAL VOLUME:
    0.2-1.1 litres
      BREATHS PER MINUTE: 20-12
      I:E RATIO 1:2
      AUTOMATIC FLOW RATE: 12-39.6 LPM
      MANUAL FLOW RATE: As per selected setting
      DEMAND FLOWRATE: Minimum 130 LPM
      MINIMUM INPUT PRESSURE: 50 PSI
      MAXIMUM INPUT PRESSURE:
    150 PSI
      MAX. AIRWAY PRESSURE:
    60 cm H2O
      OPERATING TEMPERATURE: -18°C to +50°C
    0°F to +122°F
      STORAGE TEMPERATURE: -40°C to +60°C
    -40°F to + 140°F
      PATIENT CONNECTION: 9/16" DISS
      CYLINDER DURATION: (a) 20 BPM / 0.2 I VT 100 minutes (approx.)
    (b) 12 BPM / 1.1 I V 30 minutes (approx.)
    (Based on Aluminum "D" size cylinder containing 415 litres of oxygen)
     
    (All specifications are subject to a tolerance of +/- 10%
    except the I:E Ratio which is subject to a tolerance of +/- 20%)

    Part Number and Product Description  

      01FG5000 Genesis® II A/C (Adult/Child) Handheld Automatic Time/volume Cycled Resuscitator c/w 6 foot supply hose and Universal Face Mask
     
    Genesis® II IDLH  
       
       
    Download PDF
    spec. sheet
    View product comparison chart
    Download video
     

    The specific needs of confined space/IDLH (Immediately Dangerous to Life and Health) atmosphere rescue has presented new challenges over the last few years. The IDLH Rescue Resuscitator has been developed to meet the exacting requirements of "Confined Space Rescue".

    The IDLH Rescue Resuscitator model combines simplicity of operation with technological sophistication to provide automatic ventilation, at a rate and volume designed to reduce the risk of gastric insufflation. It also provides Demand Breathing and "Automatic Circuit Shut Off" of the automatic cycling unit should the patient begin adequate spontaneous respiration.

    The IDLH Rescue Resuscitator, designed according to the specifications developed in conjunction with Shell Canada and the University of Florida, has no external controls or manual button as the confined space conditions are considered to be so hazardous that the less the operator has to do to ventilate the patient, the quicker the extrication and the greater the chance of successful resuscitation.

    An added feature of this unit is the MASK PURGE (2 lpm) CONSTANT BLEED FLOW. This reduces the risk of drawing toxic atmosphere into the mask if a leak in the mask to face seal occurs.

    The IDLH Rescue Resuscitator provides individuals with a safe and effective means of maintaining artificial ventilation during respiratory arrest, which may occur during confined space, toxic atmosphere, rescue procedures.

    Features  

  • Totally pneumatic operation
  • Lightweight
  • Durable
  • Automatic Only Cycling
  • No controls
  • Demand Breathing Capability with auto circuit shut off
  • Mask Purge Bleed Flow
  • Preset Pressure Relief
  • 2 Year Warranty Against Manufacturers Defects

    Specifications  

      TIDAL VOLUME:
    1.1 litres (Optional 0.9 litres available)
      BREATHS PER MINUTE: 12 (Optional 10 available)
      I:E RATIO: 1:2
      AUTOMATIC FLOW RATE: 40 LPM
      DEMAND FLOWRATE: Minimum 130 LPM
      MASK PURGE BLEED FLOW: 2 LPM
      MINIMUM INPUT PRESSURE: 50 PSI
      MAXIMUM INPUT PRESSURE: 150 PSI
      MAX . AIRWAY PRESSURE:
    60 cm H2O (Optional 45 cm H2O)
      OPERATING TEMPERATURE: -35°C to +60°C
    -37°F to +140°F
      STORAGE TEMPERATURE: -40°C to +60°C
    -40°F to + 140°F
      PATIENT CONNECTION: 9/16" DISS
      CYLINDER DURATION: 30 minutes (approx.)
    (Based on Aluminum "D" size cylinder containing 415 litres of oxygen)
     
    (All specifications are subject to a tolerance of +/- 10%
    except the I:E Ratio which is subject to a tolerance of +/- 20%)

    Part Number and Product Description  

      01FG7000
    Genesis® II IDLH Handheld Automatic Time/volume Cycled Rescue Resuscitator c/w 6 foot supply hose and Universal Face Mask
      01FG7100 Genesis® II IDLH Handheld Automatic Time/volume Cycled Rescue Resuscitator kit with Adjustable 0 – 25LPM Regulator with 2 DISS Self-Sealing outlets in a moulded Carrying Case c/w 6 foot supply hose and Universal Face Mask
      01FG7200 Genesis® II IDLH Handheld Automatic Time/volume Cycled Rescue Resuscitator kit with Adjustable 0 – 25LPM Regulator with 2 DISS Self-Sealing outlets in a Softpack “Clamshell” Carrying Case c/w 6 foot supply hose and Universal Face Mask
     
    Resuscitator/Ventilator Accessories  
       
       
    View product comparison chart
    Download video
     

    The CAREvent® and Genesis® II ventilator/resuscitator accessories range have been designed for use with our Automatic Transport Ventilators and resuscitators. These high quality/high performance components are an essential adjunct to our resuscitation and ventilation products.

    Because of their high quality, our Transport Ventilator Circuits can be utilized on ventilators from other manufacturers however we do not recommend the use of other manufacturer’s products on our ventilators and resuscitators.

    Features  

  • Manufactured from the finest materials
  • Durable
  • High performance
  • High quality

    Part Number and Product Description  

      01CV8001-cs
    CAREvent® BLS Single Use Patient Valve c/w Universal Facemask – case/25
      01CV8005-cs CAREvent® ALS Single Use Patient Valve with PEEP port, Universal Facemask – case/25
      01CV8010-cs CAREvent® Single Use Transport Ventilator Circuit– case/10
      01CV8015-cs CAREvent® Single Use Deluxe Transport Ventilator Circuit with PEEP port– case/10
      01CV8025 CAREvent® Deluxe Reusable Transport Ventilator Circuit with PEEP port
      01CV8050 Gurney Clip (allows handheld resuscitators to be attached to the stretcher rail)
      17MP7010 Disposable PEEP Valve case/10
      17MP7327-cs Air Intake filter for ATVs case/10
      01CV7025 CAREvent® MRI Mounting Plate Base
      01FG6500 Genesis® II Mask Retaining Insert
      01FG6501 Genesis® II Mask Retaining Insert Extraction tool
      01FG6502 Head Harness System
      01FG6520 SCBA Resuscitator Belt Pouch

     
       

    O-Two Medical Technologies Inc. reserves the right to change, modify, or update the product specifications, product data and product information contained herein at any time without prior notice.