| CAREvent® Automatic Transport Ventilators |
|

The CAREvent® Automatic Transport Ventilators are lightweight, portable, durable and simple to use, control module style, automatic time/volume cycled ventilators. They are completely pneumatic (no batteries required) and deliver breath by breath consistency in delivered tidal volume, frequency, inspiratory/expiratory times and flow rates. This consistency is not achievable with a bag-valve-mask resuscitator and prevents gastric inflation and high intrathoracic pressure.
Designed primarily for the demands of emergency rescue, resuscitation and critical care patient transports, the CAREvent® ventilators provide functionality normally only found in high end electronic ventilators.
Four models are available to meet the various functionality and budgetary needs. Features available include:
- “Demand Breathing” for the spontaneously breathing patient
- Manual Override Ventilation to supplement automatic ventilations if required and activate delayed cycling when a 30:2 CPR protocol is being performed.
- Flexibility in breath volume and frequency delivery to ventilate patients from 5 kg. to large adults (note ALS+ suitable for patients from 15 kg.)
- CPAP to provide non-invasive treatment for patients with CHF, COPD and PE. This provides one space saving device for oxygen therapy, CPAP treatment, resuscitation and ventilation (ALS+, ATV+ and MRI).
- Auto Cycling Shut-Off and Re-Start to synchronized ventilation for patients moving in and out of respiratory arrest. Note: In the child settings the inspired flow rate of the patient may be insufficient to cause the Automatic Circuit Shut off to function.
- Airway Pressure Gauge to provide visual monitoring of the patient and ventilator settings.
- An Audible and Visual Low Gas Supply Warning.
- Audible and Visual Indicator for a circuit disconnect or when delivered tidal volumes are insufficient (ATV+ and MRI)
|
| |
 |
| CAREvent® MRI |
 |
|

| |
MINUTE VOLUME: |
2 - 14 litres |
| |
BREATHS PER MINUTE: |
8 - 40 |
| |
I:E RATIO: |
1:2 |
| |
AUTOMATIC FLOW RATE: |
12- 40 LPM |
| |
MANUAL FLOW RATE: |
As per selected setting |
| |
DEMAND FLOWRATE: |
Minimum 120 LPM |
| |
DEMAND TRIGGERING PRESSURE: |
-2 cm H20 |
| |
MINIMUM INPUT PRESSURE: |
45 PSI |
| |
MAXIMUM INPUT PRESSURE: |
70 PSI |
| |
MAX . AIRWAY PRESSURE: |
60 cm H2O |
| |
OXYGEN CONCENTRATION: |
60 or 100% |
| |
ADJUSTABLE AIRWAY PRESSURE RANGE: |
20 – 60 cm H2O |
| |
OPERATING TEMPERATURE: |
-18°C to +50°C
0°F to +122°F |
| |
STORAGE TEMPERATURE: |
-40°C to +60°C
-40°F to + 140°F |
| |
PATIENT CONNECTION: |
9/16” DISS |
| |
PATIENT VALVE DEADSPACE: |
8 ml |
| |
GAS SUPPLY STATUS INDICATOR: |
VISUAL |
| |
CIRCUIT DISCONNECT ALARM: |
PULSED AUDIBLE, AND VISUAL.
ALARM SILENCE CONTROL – 15 SECONDS |
| |
LOW INPUT PRESSURE ALARM: |
PULSED VISUAL and AUDIBLE |
| |
(All specifications are subject to a tolerance of +/- 10%
except the I:E Ratio which is subject to a tolerance of +/- 20%) |
 |
Part Number and Product Description |
 |
|
 |
| |
01CV7000 |
CAREvent® MRI Control Automatic Transport Ventilator c/w 6 foot supply hose, Deluxe Disposable Transport Ventilator Circuit, Test Lung and Universal Face Mask |
| |
01CV7026 |
CAREvent® MRI Compatible Mounting Base Plate |
| |
01CV7030 |
CAREvent® ATV+ / MRI Ambulance Mounting Bracket |
|
|
|
| |
|
|